ASTM F748-25材料和器械生物试验方法选择的标准实施规程
VIP免费
1
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards,Guides and Recommendations ssued by the World Trade Organization Technical Barriers to Trade(TBT)Committee.
Designation:F748-25
INTERNATIONAL
Standard Practice for
Selecting Biological Test Methods for Materials and
Devices
This standard is issued under the fixed designation F748;the number immediately following the designation indicates the year of original
adoption or,in the case of revision,the year of last revision.A number in parentheses indicates the year of last reapproval.A superscript
epsilon(e)indicates an editorial change since the last revision or reapproval.
1.Scope
1.1 This practice describes biological test methods(as listed
in 2.1)which may be used in support of a biological evaluation
of materials and medical devices according to end-use appli-
cations.Biological evaluation is a component of a broader
safety evaluation and is conducted within an ISO 14971 risk
management framework.ISO 10993-1 sets out the general
principles and requirements for such an evaluation.A biologi-
cal evaluation conducted according to ISO 10993-1 requires a
consideration of the composition of the medical device and the
review of pre-existing information pertinent to biological risks.
In many cases this pre-existing information will be sufficient to
establish safety without a need for additional testing.Only
where existing information is insufficient it may be necessary
to conduct further testing (see ISO 10993-1).The purpose of
this document is to provide guidance on available and appli-
cable test methods to support such testing activities if they are
required.
1.2 The reader is cautioned that the area of biocompatibility
testing is a rapidly evolving field,and improved methods are
becoming available continuously,so this practice is a guide-
line.A thorough knowledge of current techniques and research
is critical to a complete evaluation of new materials and
medical devices.
1.3 Biological evaluation of materials and medical devices
for human application is discussed.Tests include those per-
formed on materials,final finished medical devices,and
extracts.The current state of the art is included for all test
procedures described.
1.4 The biocompatibility of materials used in single or
multicomponent medical devices for human use depends to a
large degree on the particular nature and duration of tissue
contact during intended use.It is not possible to specify a set
of test methods which will be necessary and sufficient to
establish biocompatibility for all materials and applications.
¹This practice is under the jurisdiction of ASTM Committee F04 on Medical and
Surgical Materials and Devices and is the direct responsibility of Subcommittee
F04.16 on Biocompatibility Test Methods.
Current edition approved Feb15,2025.Published February 2025.Originally
approved in 1982.Last previous edition approved in 2016 as F748-16.DOI:
10.1520/F0748-25.
Because biological reactions caused by a material or a medical
device can have different implications or raise different safety
concerns depending on the type and duration of tissue contact
of the medical device,each application must be evaluated for
biocompatibility and any associated testing on its own merit.
1.5 The evaluation of tissue-engineered medical products
(TEMPs)may,in some cases,involve different or additional
testing beyond that conducted on non-TEMPs-based materials
and medical devices.Where appropriate,these differences are
discussed in this practice and additional tests described.For
additional information,see reference standards published un-
der the jurisdiction of the applicable TEMPs Subcommittees of
F04.
1.6 The ethical use of research animals places the obligation
on the individual investigator to determine the most efficient
methods for performing the necessary testing without undue
use of animals.Where adequate prior data exists to substantiate
certain types of safety information,these guidelines should not
be interpreted to mean that testing should be unnecessarily
repeated.Refer to ISO 10993-2 for additional information.
1.7 This standard does not purport to address all of the
safety concerns,if any,associated with its use.It is the
responsibility of the user of this standard to establish appro-
priate safety,health,and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
1.8 This international standard was developed in accor-
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
Development of International Standards,Guides and Recom-
mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT)Committee.
2.Referenced Documents
2.1 ASTM Standards:²
F619 Practice for Extraction of Materials Used in Medical
Devices
²For referenced ASTM standards,visit the ASTM website,www.astm.org,or
contact ASTM Customer Service at www.astm.org/contact.For Annual Book of
ASTM Standards volume information,refer to the standard's Document Summary
page on the ASTM website.
Copyright ◎ASTM Intermational,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United States
2
F748-25
F719 Practice for Testing Materials in Rabbits for Primary
Skin Irritation
F720 Test Method for Testing Guinea Pigs for Contact
Allergens:Guinea Pig Maximization Test
F749 Practice for Evaluating Material Extracts by Intracuta-
neous Injection in the Rabbit
F750 Practice for Evaluating Acute Systemic Toxicity of
Material Extracts by Systemic Injection in the Mouse
F756 Practice for Assessment of Hemolytic Properties of
Materials
F763 Practice for Short-Term Intramuscular Screening of
Implantable Medical Device Materials
F813 Practice for Direct Contact Cell Culture Evaluation of
Materials for Medical Devices
F895 Test Method for Agar Diffusion CellCulture Screening
for Cytotoxicity
F981 Practice for Assessment of Muscle and Bone Tissue
Responses to Long-Term Implantable Materials Used in
Medical Devices
F1408 Practice for Subcutaneous Screening Test for Implant
Materials
F1439 Guide for Performance of Lifetime Bioassay for the
Tumorigenic Potential of Implant Materials
F1841 Practice for Assessment of Hemolysis in Continuous
Flow Blood Pumps
F1877 Practice for Characterization of Particles
F1903 Practice for Testing for Cellular Responses to Par-
ticles in vitro
F1904 Guide for Testing the Biological Responses to Medi-
cal Device Particulate Debris and Degradation Products in
vivo
F1983 Practice for Assessment of Selected Tissue Effects of
Absorbable Biomaterials for Implant Applications
F2147 Practice for Guinea Pig:Split Adjuvant and Closed
Patch Testing for Contact Allergens
F2148 Practice for Evaluation of Delayed Contact Hyper-
sensitivity Using the Murine Local Lymph Node Assay
(LLNA)
F2382 Test Method for Assessment of Circulating Blood-
Contacting Medical Device Materials on Partial Throm-
boplastin Time(PTT)
F2888 Practice for Platelet Leukocyte Count—An In-Vitro
Measure for Hemocompatibility Assessment of Cardio-
vascular Materials
F2901 Guide for Selecting Tests to Evaluate Potential Neu-
rotoxicity of Medical Devices
2.2 Other Referenced Documents:
ISO 7405 Evaluation of biocompatibility of medical devices
used in dentistry³
ISO 10993-1 Biological evaluation of medical devices—Part
1:Evaluation and testing within a risk management
process³
ISO 10993-2 Biological evaluation of medical devices—Part
2:Animal welfare requirements³
ISO 10993-3 Biological evaluation of medical devices—Part
³Available from American National Standards Insitute(ANSD),25W.43rd St.,
4th Floor,New York,NY 10036,http://www.ansi.org.
3:Tests for genotoxicity,carcinogenicity and reproductive
toxicity³
ISO 10993-4 Biological evaluation of medical devices—Part
4:Selection of tests for interactions with blood³
ISO 10993-5 Biological evaluation of medical devices—Part
5:Tests for in vitro cytotoxicity³
ISO 10993-6 Biological evaluation of medical devices—Part
6:Tests for local effects after implantation³
ISO 10993-10 Biological evaluation of medical devices—
Part 10:Tests for skin sensitization³
ISO 10993-11 Biological evaluation of medical devices—
Part 11:Tests for systemic toxicity³
ISO 10993-12 Biological evaluation of medical devices—
Part 12:Sample preparation and reference materials³
ISO 10993-20 Biological evaluation of medical devices—
Part 20:Principles and methods for immunotoxicology
testing of medical devices³
ISO 10993-22 Biological evaluation of medical devices—
Part 22:Guidance on nanomaterials³
ISO 10993-23 Biological evaluation of medical devices—
Part 23:Tests for irritation³
ISO 14971 Medical devices—Application of risk manage-
ment to medical devices³
ISO 18562-1 Biocompatibility evaluation of breathing gas
pathways in healthcare applications—Part 1:Evaluation
and testing within a risk management process³
ISO 22442 Medical devices utilizing animal tissues and their
derivatives ³
ISO/TS 37137-1 Biological evaluation of absorbable medi-
cal devices—Part 1:General requirements³
ANSI/AAMI ST72 Bacterial endotoxins—Test methods,
routine monitoring,and alternatives to batch testing³
USP<85> Bacterial Endotoxins Test⁴
USP<151> Pyrogen Test⁴
USP<161> Medical Devices—Bacterial Endotoxin and Py-
rogen Tests⁴
FDA Use of International Standard ISO 10993-1,“Biologi-
cal evaluation of medical devices-Part 1:Evaluation and
testing within a risk management process”-Guidance for
Industry and Food and Drug Administration Staff(Sep-
tember 8,2023)⁵
FDA Immunotoxicity Testing Guidance⁶
FDA Guidance for Industry Preparation of IDEs and INDs
for Products Intended to Repair or Replace Knee Carti-
lage⁷
FDA Knee Joint Patellofemorotibial and Femorotibial
Metal/Polymer Porous-Coated Uncemented Prostheses-
Class II Special Controls Guidance Document for Industry
and FDA⁸
4Available from U.S.Pharmacopeial Convention(USP),12601 Twinbrook
Pkwy.,Rockvile,MD 20852-1790,http://www.usp.org.
5https://www.fda.gov/media/142959/download
⁶https://www.fda.gov/regulatory-information/search-fda-guidance-documents/
immunotoxicity-testing-guidance
7https://www.fda.gov/media/82562/download
8https://www.fda.gov/medical-devices/guidance-documents-medical-devices-
and-radiation-emitting-products/knee-joint-patellofemorotibial-and-femorotibial-
metalpolymer-porous-coated-uncemented-prostheses
F748-25
3
3.Terminology
3.1 Definitions:
3.1.1 endpoint—the term refers to certain parameters mea-
sured as primary or surrogate outcomes for biological or toxic
responses under assessment.
3.1.2 novel material—the term refers to a material that has
not been previously used in any legally marketed device.9
4.Significance and Use
4.1 The objective of this practice is to provide information
to consider when selecting and conducting biological testing,
when deemed necessary.
4.2 A matrix listing the biological endpoints relevant to a
biocompatibility evaluation of materials,medical devices,and
their applications is included in Table 1.The expected nature of
tissue contact and duration of use of a medical device is also
included.The position of row and column intersection is
marked to indicate the types of biological tests that should be
considered for a material or medical device for the specific
application,when testing is warranted.
5.Classification of Materials and Devices by End-Use
Applications
5.1 General:
5.1.1 Classification should focus on device materials that
will have direct or indirect tissue contact.Manufacturing
processes that can potentially alter the chemical,physical,or
biological characteristics of a raw material should be consid-
ered.It is preferable to obtain material test articles directly
from the final medical device to ensure that the manufacturing
processes,including cleaning,packaging,sterilization,and any
other procedures used in production and preparation of the
medical device are evaluated.Consideration should be given to
reprocessed medical devices and the effects reprocessing may
have on the biocompatibility profile.Novel materials may
require specialized testing to ensure they do not impart
unacceptable risks if not appropriately characterized.
5.1.2 In addition to considering the duration and type of
tissue contact of a candidate material or medical device,use of
a scaling factor is an important consideration when determin-
ing the relative size and/or surface area of sampled material
needed to exaggerate human dose/exposure for certain end-
points(e.g.,systemic toxicity,implantation,carcinogenicity).
The material surface area to body weight ratio and other
proportional parameters may be especially important for test-
ing porous materials and medical devices with repeated appli-
cations(e.g.,dialysis products).These also could be important
when assessing the potential risks of exposure to substances
released from a material over time,or exposure to degradation
products.
5.1.3 The first step in biological evaluation is to categorize
the medical device based on its intended tissue contact type
(refer to 5.2)and duration,which helps to facilitate the
9 The term “Iegally marketed device”and the necessary supporting biocompat-
iblity information may vary depending on the regulatory jurisdiction.
selection of relevant biological endpoints.The contact duration
can be divided into the following categories:
(1)Transitory contact(<1 min):very brief contact with the
body(e.g.,lancets,hypodermic needles,capillary tubes that are
used for less than one minute).Due to very brief contact time,
the risks in general are considered to be low,and testing may
not be needed.Note—Other terminology may exist for this
type of contact.
(2)Limited contact(≤24 h).
(3)Prolonged contact(24 h to 30 days).
(4)Long-term contact(>30 days).
5.1.3.1 When considering duration of contact for devices
that are intended for repeat use(e.g.,reusable devices or
single-use devices applied more than once to the patient during
treatment),cumulative exposure should also be considered and
applied to calculate the overall contact duration for use on a
single patient.The contact duration shall be based on the repeat
or cumulative use because,if chemicals are leached from the
device,those chemicals can bioaccumulate such that with
repeat medical device use,higher levels of chemicals may be
present in amounts that could result in adverse biological
response.Therefore,when determining the contact duration
categorization of repeat use devices,the total exposure period
in days of use of the medical device(e.g.,one medical device
can be reused intermittently over a number of days,or
replacement devices may be used over a number of days)on a
single patient should be considered,regardless of the duration
of exposure on each day.Cumulative or repeat use examples:
(1)Limited contact(≤24 h):
(A)A medical device used for three sessions within a
period of 24 h,20 min per session.
(B)A medical device used for 5 min,once every 30 days,
for a period of six months and there is specific information that
supports that bioaccumulation is unlikely(e.g.,elimination
time is less than one day).When utilizing this approach,it is
recommended to discuss the relevance of the available infor-
mation with the applicable regulatory body to determine the
appropriate contact duration.
(2)Prolonged contact(24 h to 30 days):
(A)A medical device used 20 min each day for five days.
(B)A medical device used for 5 min,once every 30 days,
for a period of six months and there is specific information that
supports that bioaccumulation is unlikely(e.g.,elimination
time is less than 30 days).When utilizing this approach,it is
recommended to discuss the relevance of the available infor-
mation with the applicable regulatory body to determine the
appropriate contact duration.
(3)Long-term contact(>30 days):A medical device used
for 5 min once a day for two months and information regarding
bioaccumulation is not available.
5.1.4 In general,the assessment for medical devices con-
taining biologically derived materials (e.g.,collagen based,
demineralized bone matrix,hyaluronic acid)or biologics
should address the same considerations specific to the type,
location,and duration of tissue contact as other medical
devices and products.Additional assessments for TEMPs are
provided in other standards (e.g.,ISO 22442)and may be
considered as part of an overall testing strategy.
摘要:
展开>>
收起<<
1ThisinternationalstandardwasdevelopedinaccordancewithinternationallyrecognizedprinciplesonstandardizationestablishedintheDecisiononPrinciplesfortheDevelopmentofInternationalStandards,GuidesandRecommendationsssuedbytheWorldTradeOrganizationTechnicalBarrierstoTrade(TBT)Committee.Designation:F748-25IN...
声明:如果您的权利被侵害,请联系我们的进行举报。
相关推荐
-
RBA8.0官方中文版VIP免费
2025-11-16 19 -
新版RBA责任商业联盟一整套文件(含管理手册、全套程序文件及对应表单)
2025-11-17 35 -
最新最全面RBA(原EICC)社会责任程序文件(77个程序+相关表格)
2025-11-17 28 -
RBA-SP-067 供应商分包商实施RBA管理程序
2025-11-18 25 -
RBA-SP-066 记录管理程序
2025-11-18 25 -
RBA-SP-065 文件管理程序
2025-11-18 20 -
RBA-SP-064 纠正和预防措施管理程序RBA8.0
2025-11-18 24 -
RBA-SP-063 管理评审控制程序
2025-11-18 19 -
RBA-SP-062 内部审核管理程序
2025-11-18 29 -
RBA-SP-006 劳务派遣管理程序
2025-11-18 29
作者:冒牌货
分类:法规规范
价格:80质量币
属性:9 页
大小:1.35MB
格式:PDF
时间:2025-12-31

