ASTM F2382-24 循环血液接触医疗器械材料对部分凝血活酶时间(PTT)的评估
VIP免费
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards,Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade(TBT)Committee.
Designation:F2382-24
INTERNATIONAL
Standard Test Method for
Assessment of Circulating Blood-Contacting Medical Device
Materials on Partial Thromboplastin Time(PTT)¹
This standard is issued under the fixed designation F2382;the number immediately following the designation indicates the year of
original adoption or,in the case of revision,the year of last revision.A number in parentheses indicates the year of last reapproval.A
superscript epsilon(ε)indicates an editorial change since the last revision or reapproval.
1.Scope
1.1 This test method covers the screening of circulating
blood-contacting device materials for their ability to induce
blood coagulation via the intrinsic coagulation pathway.This
assay should be part of the hemocompatibility evaluation for
devices and materials contacting human blood,as per ANSI/
AAMIISO 10993-4.See also Practice F2888.
1.2 All safety policies and practices shall be observed
during the performance of this test method.
1.3 All plasma and any materials that had contact with
plasma will be bagged in a biohazard bag,properly labeled
with the contents,and disposed of by appropriate means.The
plasma should be handled at the Biosafety Level 2(BSL-2)as
recommended in the Centers for Disease Control(CDC)
National Institutes of Health(NIH)Manual Biosafety in
Microbiological and Biomedical Laboratories(BMBL,current
edition).
1.4 The normal pooled human plasma must have tested
negative for Hepatitis B(HBV)and Human Immunodeficiency
(HIV)viruses.The plasmas should be treated like any patient
plasma using standard precautions.The plasma should be
handled at the BSL-2 as recommended in the CDC/NIH
Manual BMBL.
1.5 The values stated in SI units are to be regarded as
standard.No other units of measurement are included in this
standard.
1.6 This standard does not purport to address all of the
safety concerns,if any,associated with its use.It is the
responsibility of the user of this standard to establish appro-
priate safety,health,and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accor-
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
IThis test method is under the jurisdiction of ASTM Committe F04 on Medical
and Surgical Materials and Devices and is the direct responsibility of Subcommittee
F04.16 on Biocompatibility Test Methods.
Current edition approved Aug.15,2024.Published August 2024.Originally
approved in 2004.Last previous edition approved n 2018 as F2382-18.DOI:
10.1520/F2382-24.
Development of International Standards,Guides and Recom-
mendations issued by the World Trade Organization Technical
Barriers to Trade(TBT)Committee.
2.Referenced Documents
2.1 ASTM Standards:²
F2888 Practice for Platelet Leukocyte Count—An In-Vitro
Measure for Hemocompatibility Assessment of Cardio-
vascular Materials
2.2 ANSI/AAMI Standard:3
ANSI/AAMI/ISO 10993-4 Biological Evaluation of Medical
Devices—Part 4:Selection of Tests for Interactions with
Blood
2.3 Other Document:
U.S.Department of Health and Human Services Biosafety in
Microbiological and Biomedical Laboratories(BMBL),
5th ed.,1999⁴
3.Terminology
3.1 Definitions of Terms Specific to This Standard:
3.1.1 activator—a medical material which demonstrates a
shortened clotting time;an initiator of the intrinsic coagulation
pathway.
3.1.2 blank time—a period at the beginning of an assay
when no data is taken.This is done to eliminate interference
from premixing reagents,bubbles,and so forth.
3.1.3 duplicate flag—the agreement between the results of
duplicate samples in percent.For example,if set to "15,"the
difference between the two channels must be less than or equal
to 15%.If the variance in clot times exceeds this percentage,
an asterisk“*”will be printed by the average results on the
report.
²For referenced ASTM standards,visit the ASTM website,www.astm.org,or
contact ASTM Customer Service at www.astm.org/contact.For Annual Book of
ASTM Standards volume information,refer to the standard's Document Summary
page on the ASTM website.
³Available from American National Standards Institute(ANSI),25W.43rd St.,
4th Floor,New York,NY 10036,http://www.ansi.org.
4The BMBL5th Edition (December 2009)is available from the Government
Printing Office or https://www.cdc.gov/biosafety/publications/bmbl5/bmbl.pdf
Copyright ◎ASTM Intermational,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United States
Eqr{tkijvd{"CUVO"pin"cm'tkijw'tgugtxgf+'Uer'Qev27"365262'WVE'4246 1
F2382-24
3.1.4 equilibration time—the time allowed for the plasma
samples to warm to 37℃.The coagulation analyzer can be set
to zero if samples are pre-warmed to this temperature.
3.1.5 partial thromboplastin time(PTT)assay—a modifica-
tion of the Activated Partial Thromboplastin Time(APTT)
assay;unlike the APTT test,the PTT assay uses a reagent
(rabbit brain cephalin)without activating substances such as
silica,kaolin,elagic acid.The material being tested acts as the
activator.
3.1.6 read time—the time during which data is collected to
detect a clot.
4.Significance and Use
4.1 The purpose of this test method is to determine the time
citrated plasma exposed to medical materials takes to form a
clot when exposed to a suspension of phospholipid particles
and calcium chloride.In this test method,the test article is the
activator.The PTT assay is a general screening test for a
medical material's ability to activate the intrinsic coagulation
pathway.Material samples that show a shortened PTT are
activators of the intrinsic coagulation pathway.
4.2 The test article,reference materials,and controls are
exposed to human plasma.The plasma is tested on a coagula-
tion device.Each sample tube is assayed in duplicate.The
results are reported as a percentage of the negative control.
5.Apparatus
5.1 Polypropylene Test Tubes with Caps,12 by 75 mm.
5.2 Automatic Pipets and Tips,100 and 1000 μL.
5.3 Ice Bath.
5.4 Coagulation Analyzer;qualified,see A1.1.
5.5 Agitating Water Bath,37±2℃,capable of 60 rpm.
5.6 Coagulation Analyzer Cuvettes,or equivalent for spe-
cific analyzer.
6.Reagents and Materials
6.1 Calcium Chloride,25 mM.
6.2 Citrated Human Blood Plasma,fresh (less than 4h from
draw)or freshly frozen,maintained at minus 80℃,pooled
from at least three donors.
7.Hazards
7.1 The human blood plasma should be treated like any
patient plasma using standard precautions.The plasma should
be handled as recommended for BSL-2 by the CDC/NIH
publication Biosafety in Microbiological and Biomedical
Laboratories(BMBL,current edition).
8.Preparation of Apparatus
8.1 Prepare each test article,the negative reference
materials,marketed comparator device (if used),and controls
in triplicate.If a positive reference control material is used,a
single replicate is acceptable.All samples are prepared based
on a ratio of either 4 or,preferably 6 cm²of material to 1 mL
plasma and placed into polypropylene tubes.For device
testing,if test sample quantity allows,use three separate
devices;otherwise,take three representative samples from one
device.
8.2 Label duplicate polypropylene tubes and place in the ice
bath.
8.3 Turn on the coagulation analyzer and allow it to warm
up to 37±2℃ and equilibrate for at least 10 min.
8.4 Program the analyzer to test under the APTT function
with an equilibration time of 60 s,activation time of 120 s,a
read time of at least 300 s (that is,long enough to see plasma
clot formation),and an analyzer-specific blank time (if appli-
cable).
8.5 Pre-warm analyzer cuvettes(or cups,depending on
analyzer selected)at 37±2℃.
8.6 Pre-warm calcium chloride at 37±2℃.
8.7 Rabbit Brain Cephalin(RBC)Preparation:
8.7.1 Allow the RBC to come to room temperature.
8.7.2 Reconstitute the RBC as indicated by the RBC reagent
manufacturer.
8.7.3 Place in an agitating water bath set at 37±2℃ and
60 rpm for 15 min to ensure complete rehydration of contents.
8.7.4 Vortex 15 s after rehydration is complete.
8.7.5 Place at 37±2℃.
8.8 If using frozen blood plasma,quick thaw the plasma at
37±2℃ and place on ice immediately.
6.3 Lyophilized Rabbit Brain Cephalin(RBC).
6.4 Positive Reference Material(Optional),see A1.2.
6.5 Positive Control,glass(Pasteur pipette tips or glass
beads).Other qualified positive control materials such as
Buna-N-Rubber may be selected once they have demonstrated
a consistent thrombogenic response.
6.6 Negative Reference Material (for example,high-density
polyethylene,HDPE).
6.7 Marketed Comparator Device(Optional)—A legally
marketed,clinically acceptable device that has similar blood
contact nature and clinical use as the material/device being
investigated.
NoTE 1—It may be helpful to use a positive reference control material
(n=1)per assay to ensure continuity between runs.
9.Procedure
9.1 The test material(s),reference material(s),marketed
comparator device (if used),and controls are placed into
polypropylene tubes and exposed to the appropriate quantity of
plasma,based on a ratio of 4 cm²,or preferably 6 cm²of
material to 1 mL plasma.The negative control is a polypro-
pylene tube with 1 mL of plasma,without additional material.
9.2 The samples are exposed to the plasma for 15±1 min
in a 37±2℃ agitating water bath at 60 rpm.
9.3 After 15 min of incubation,the tubes are immediately
placed into the ice bath and immediately transferred into
pre-chilled new polypropylene tubes.
9.4 Vortex each sample 15 s before each use/run.
Eqr{tkijvd{"CUVO"pin"cm'tkijw'tgugtxgf+'Uer'Qev27"365262'WVE'4246 2
Eqr{tkijvd{"CUVO"pin"cm'tkijw'tgugtxgf+'Uer'Qev27"365262'WVE'4246 3
9.5 Avoiding bubbles,transfer 100 μL of the plasma into
pre-warmed cuvettes and allow the plasma to equilibrate for
60s at 37±2℃.
9.6 To each cuvette/cup,add 100 μL warmed RBC
preparation,initiating the 2 min activation step.(Invert RBC to
mix prior to each use.)
9.7 After activation,add 100 μL warmed 25 mM calcium
chloride to each cuvette.
9.8 Allow the analyzer to read the sample for the formation
of clots(up to 5 min).
9.9 Record the clotting time(seconds)for each sample,as
well as the average clotting time of the duplicate samples.
NoTE 2—The volume ofplasma,RBC reagent,and calcium chloride
needed for the clotting time measurement may vary with different
anticoagulation analyzers.It is acceptable to use a plasma volume that is
specific to the coagulation analyzer used,as long as the plasma to reagent
ratio remains 1:1.
10.Calculation of Results
10.1 For each replicate,calculate the average clotting time
(from two readings)of the test sample,negative reference
control material,positive control material,and marketed com-
parator device,normalized to the negative control (untreated
plasma)values.In each case,the average clotting time is
expressed as a percentage relative to the negative control.
%negative control= (1)
10.2 Statistical Analysis of Results:
10.2.1 Calculate the mean values and standard deviations
for the test sample,negative control reference material,posi-
tive control material,and marketed comparator device by
averaging the respective percentage values from all three
replicates.
10.2.2 The positive control and the positive reference ma-
terial control (if used)results,expressed as percent negative
control(untreated plasma),must generally be<50%.However,
positive control results between 50 to 60%that of the negative
control may also be acceptable if the difference between the
negative reference control and the positive control is at least
30%(for example,if the result for the positive control=55%,
then the value for the negative reference control should be
>85%).The negative reference material control result,ex-
pressed as percent negative control(untreated plasma),must be
>80 %.If the assay does not meet this specification,the
experiment is to be repeated until the controls meet these
specifications.
10.2.3 The coefficient of variation(CV)between the dupli-
cates for each sample must be ≤15%.The duplicates of each
test article sample are averaged,and one value is reported as
the clotting time.This results in three clotting time values for
each test article.The three values are then averaged to report a
final average clotting time of the test article.The values for
each test article sample must be within±25%of this average.
If the values are greater than 25%of the average of the run,the
experiment needs to be repeated.
10.2.4 The values for the test article,negative control,
positive control,negative and/or positive reference material,
marketed comparator device results are compared to each other
with analysis of variance(ANOVA)including post-hoc pair-
wise comparisons such as a Tukey or Newman-Keuls test.
10.2.5 Differences shall be considered statistically signifi-
cant if the p-value is less than 0.05.The test article is
considered to pass the PTT test if there is no statistical
difference between the test article and the negative control
(untreated plasma)or the negative reference material control.
11.Assay Validity
11.1 The positive control and the positive reference material
control(if used)results,expressed as percent negative control
(untreated plasma),must generally be <50 %.However,posi-
tive control results between 50 and 60%of that of the negative
control may also be acceptable if the difference between the
negative and the positive control is at least 30%(for example,
if the result for the positive control=55%,then the value for
the negative reference control should be >85%).The negative
reference material control result,expressed as a percent nega-
tive control (untreated plasma),must be >80%.If the assay
does not meet this specification,this assay is to be repeated
until the controls meet these specifications.
11.2 The coefficient of variation(CV)between the dupli-
cates for each sample must be ≤15%.The duplicates of each
test article sample or legally marketed comparator device (if
used)sample are averaged and one value is reported as the
clotting time.This results in three clotting time values for each
test article and legally marketed comparator device (if used).
The three values are then averaged to report a final average
clotting time of the test article or legally marketed comparator
(if used).The values for each test article sample must be within
±25%of this average.If the values are greater than 25%of
the average of the run,the experiment needs to be repeated.
12.Test Evaluation
12.1 The test article is also considered to pass the PTT test
if the test article results are>80%(acceptance criterion for the
negative reference material)and the test article is at least 30%
above that of the positive control.If the results of the test
article are≤50%(acceptance criterion for a positive control),
then the test article is considered to fail this screening test.
12.2 If the test article results fall between these assay
acceptance criteria of 50 to 80%of the negative control,it is
recommended that the results from the test article be compared
to results from a legally marketed comparator device(that is,in
a repeat test with all controls).If the results for the test article
are less than that of a comparator,then additional thromboge-
nicity evaluation is recommended(for example,risk
assessment,animal study,or other thrombogenicity tests,or a
combination thereof,on the finished device under more clini-
cally relevant conditions)
13.Precision and Bias
13.1 The precision and bias determination of this test
method is not relevant because this is a biological assay(not an
摘要:
展开>>
收起<<
ThisinternationalstandardwasdevelopedinaccordancewithinternationallyrecognizedprinciplesonstandardizationestablishedintheDecisiononPrinciplesfortheDevelopmentofInternationalStandards,GuidesandRecommendationsissuedbytheWorldTradeOrganizationTechnicalBarrierstoTrade(TBT)Committee.Designation:F2382-24I...
声明:如果您的权利被侵害,请联系我们的进行举报。
相关推荐
-
ASTM F3172-15 血管内医疗器械设计验证数量和样本容量选择指南
2024-04-15 138 -
ASTM-D-882-测量塑料薄膜和薄片材拉伸性能(
2024-04-16 103 -
ASTM F838-2015 确定用于液体过滤的膜过滤器的细菌滞留的标准试验方法
2024-05-09 98 -
ASTM F3064-21 Standard Specification for Normal Category Aeroplanes CertificationVIP免费
2024-05-09 58 -
ASTM D3195 D3195M-10(2015) Standard Practice for Rotameter CalibrationVIP免费
2024-05-09 66 -
ASTM D4169-22 运输集装箱和系统的性能测试1(中文版)VIP免费
2025-09-09 56 -
ASTM D4169 运输集装箱和设备性能试验的标准实施规范(中英文)VIP免费
2025-09-09 42 -
ASTME2500 确证方法--关键质量属性关键工艺参数关键方面VIP免费
2025-09-26 26 -
ASTM E2500-25 药品的规格、设计与验证以及生物制药生产系统装备科学与基于风险的方法1(中文)VIP免费
2025-11-21 86 -
ASTM E3219-25 基于健康的暴露限值(HBEL)推导指南(中文)
2025-12-17 82
作者:冒牌货
分类:法规规范
价格:80质量币
属性:4 页
大小:290.24KB
格式:PDF
时间:2026-01-05

