ASTM F1877-24 Characterization of Particles

VIP免费
3.0 冒牌货 2026-01-05 10 1.12MB 16 页 100质量币
侵权投诉
1
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards,Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade(TBT)Committee.
Designation:F1877-24
INTERNATIONAL
Standard Practice for
Characterization of Particles¹
This standard is issued under the fixed designation F1877;the number immediately following the designation indicates the year of
original adoption or,in the case of revision,the year of last revision.A number in parentheses indicates the year of last reapproval.A
superscript epsilon(e)indicates an editorial change since thelast revision or reapproval.
1.Scope
1.1 This practice covers a series of recommendations,gen-
erally applicable to all medical devices,for characterization of
the morphology,shape,size,and size distribution of particles.
The methods utilized include sieves,optical,scanning electron
microscopy(SEM),transmission electron microscopy(TEM),
and electrooptical.
1.2 While characterizing the quantity or number of particles
shed from medical devices is important,this is not covered
within the scope of the current document.AAMI TIR 42 and
USP<788>provide guidelines for determination of particle
quantities in various size ranges.
1.3 These methods are appropriate for particles produced by
a number of different methods.These methods can include
simulated use approaches such as in vitro wear test machines
(Test Method F732),total joint simulation systems(Guides
F1714 and F1715),abrasion testing,and vascular durability
testing(Guide F2942).Other methods for producing particles
such as shatter boxes or pulverizers,as well as commercially
available particles,and particles harvested from tissues in
animal or clinical studies can be used.
1.4 Except for chemical composition,this standard does not
address sample preparation procedures and/or test systems that
can be affected by chemical properties(for example,solubility,
miscibility).While this standard does not provide detailed
recommendations regarding assessment of chemical properties
of particles,these should be considered.
1.5 The particles may be metallic,polymeric,or ceramic
and are released from medical device materials either acutely
or chronically(for example,due to wear).
1.6 The digestion procedures to be used and issues of
sterilization of retrieved particles are not the subject of this
practice.
1.7 A classification scheme for description of particle mor-
phology is included in Appendix X3.
!This practice is under the jurisdiction of ASTM Committe F04 on Medical and
Surgical Materials and Devices and is the direct responsibility of Subcommittee
F04.16 on Biocompatibility Test Methods.
Current edition approved Oct.1,2024.Published October 2024.Originally
approved in 1998.Last previous edition approved n 2016 as F1877-16.DOI:
10.1520/F1877-24.
1.8 When nanoparticles (that is,having at least one dimen-
sion less than 100 nm)are known to be present or are expected,
other characterization methods may be needed.For informa-
tion regarding nanoparticle characterization,refer to standards
that address nanoparticles(for example,ISO 21363,ISO/TR
10993-22,ISO/TR 16196).
1.9 This standard does not address ions released from
medical devices.
1.10 The values stated in SI units,including units officially
accepted for use with SI,are to be regarded as standard.No
other systems of measurement are included in this standard.
1.11 As a precautionary safety measure for handling test
samples during particle characterization analyses,removed
particles from implant tissues should be sterilized or minimally
disinfected by an appropriate means that does not adversely
affect these particles.
1.12 This standard does not purport to address all of the
safety concerns,if any,associated with its use.It is the
responsibility of the user of this standard to establish appro-
priate safety,health,and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
1.13 This international standard was developed in accor-
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
Development of International Standards,Guides and Recom-
mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT)Committee.
2.Referenced Documents
2.1 ASTM Standards:²
C678 Test Method for Determination of Particle Size Distri-
bution of Alumina or Quartz Using Centrifugal Sedimen-
tation (Withdrawn 1995)³
E11 Specification for Woven Wire Test Sieve Cloth and Test
Sieves
E161 Specification for Electroformed Material and Test
Sieves
²For referenced ASTM standards,visit the ASTM website,www.astm.org,or
contact ASTM Customer Service at www.astm.org/contact.For Annual Book of
ASTM Standards volume information,refer to the standard's Document Summary
page on the ASTM website.
3The last approved version of this historical standard is referenced on
www.astm.org.
Copyright ASTM Intermational,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United States
2
F1877-24
E766 Practice for Calibrating the Magnification of a Scan-
ning Electron Microscope
E1617 Practice for Reporting Particle Size Characterization
Data
F561 Practice for Retrieval and Analysis of Medical
Devices,and Associated Tissues and Fluids
F660 Practice for Comparing Particle Size in the Use of
Alternative Types of Particle Counters
F661 Practice for Particle Count and Size Distribution Mea-
surement in Batch Samples for Filter Evaluation Using an
Optical Particle Counter (Withdrawn 2000
F662 Test Method for Measurement of Particle Count and
Size Distribution in Batch Samples for Filter Evaluation
Using an Electrical Resistance Particle Counter (With-
drawn 2002
F732 Test Method for Wear Testing of Polymeric Materials
Used in Total Joint Prostheses
F1714 Guide for Gravimetric Wear Assessment of Prosthetic
Hip Designs in Simulator Devices
F1715 Guide for Wear Assessment of Prosthetic Knee De-
signs in Simulator Devices (Withdrawn 2006
F2942 Guide for in vitro Axial,Bending,and Torsional
Durability Testing of Vascular Stents
2.2 ISO Standards:⁴
ISO 21363 Nanotechnologies—Measurements of particle
size and shape distributions by Transmission Electron
Microscopy
ISO 13322-1 Particle size analysis—Image analysis
methods—Part 1:Static image analysis methods
ISO/TR 10993-22 Biological evaluation of medical
devices—Part 22:Guidance on nanomaterials
ISO/TR 16196 Nanotechnologies—Compilation and de-
scription of sample preparation and dosing methods for
engineered and manufactured nanomaterials
2.3 AAMI Document:
AAMI TIR 42:2021 Evaluation of Particulate Associated
with Vascular Medical Devices
2.4 U.S.Pharmacopeia:⁶
USP<788> Particulate Matter in Injections⁷
3.Terminology
3.1 Definitions of Terms Specific to This Standard:
3.1.1 agglomerate,n—a jumbled mass or collection of two
or more particles or aggregates,or a combination thereof,held
together by relatively weak cohesive forces caused by weak
chemical bonding or an electrostatic surface charge generated
by handling or processing.
4Available from American National Standards Institute(ANSD),1180 Avenue of
the Americas,10th Floor,New York,NY 10036,http://www.ansi.org.
5Available from Association for the Advancement of Medical Instrumentation
(AAMI),4301 N.Fairfax Dr.,Suite 301,Arlington,VA 22203-1633,http://
www.aami.org.
⁶Available from U.SPharmacopeial Convention(USP),12601 Twinbrook
Pkwy.,Rockville,MD 20852-1790,http://www.usp.org.
7USP<788>is intended to provide recommendations for particulate evaluations
of injections and parenteral infusions,and is generally not applicable to medical
devices.While limitations of this standard exist and the method used should be
validated,it provides some useful information(for example,particle quantification
methods,system cleanliness recommendations)
3.1.2 aggregate,n—a dense mass of particles held together
by strong intermolecular or atomic cohesive forces that is
stable with normal mixing techniques,including high-speed
stirring and ultrasonics.
3.1.3 aspect ratio (AR),n—a ratio of the major to the minor
Feret diameters of a particle(see 12.3.3).
3.1.4 elongation(E),n—ratio of the particle length to the
average particle width(see 12.3.4).
3.1.5 equivalent circle diameter(ECD),n—the diameter of
a circle with an area equivalent to the area of the particle (see
12.3.2 and Appendix X1).
3.1.6 Feret diameter;n—the mean value of the distance
between pairs of parallel tangents to a projected outline of a
particle.
3.1.7 form factor(FF),n—a dimensionless number relating
area and perimeter of a particle,as determined in 12.3.6.
3.1.8 irregular;,adj—referring to a particle that cannot be
described as round or spherical.A set of standard nomenclature
and reference figures are given in Appendix X2.
3.1.9 matrix,n—particle-free sample that represents the
starting composition of test samples prior to preparation (for
example,digestion/harvest)or analysis.
3.1.10 method qualification,nthe process of evaluating
the accuracy,precision,and bias of sample preparation and
particle characterization analyses of test particle suspensions
based on comparison with defined samples of particles with
similar properties.The purpose is to qualify the acceptability of
the analytical outcomes.
3.1.11 morphology,n—referring to size,shape,structure,
and texture.
3.1.12 particle,n—the smallest discrete unit detectable as
determined in test methods.
3.1.13 particle breadth,n—distance between touch points of
the shortest Feret pair,orthogonal to length.
3.1.14 particle length,n—distance between the touch points
of maximum Feret pair.This value will be greater than or equal
to the maximum Feret diameter.
3.1.15 reference particles,n—particle samples with defined
properties(for example,size,shape,composition,density,
population distribution,agglomeration)that are used for
method qualification.
3.1.16 rectangular;adj—referring to a particle that approxi-
mates a square or rectangle in shape.
3.1.17 roundness(R),n—a measure of how closely an
object represents a circle as determined in 12.3.5.
3.1.18 spherical,adj—referring to a particle that approxi-
mates a sphere in a three-dimensional space and that projects as
a round (or circular)shape in a two-dimensional space.A
sphere is a three-dimensional,geometrical shape that has all its
surface points equidistant from a common point.
3.1.19 spiked matrix,n—matrix with reference particles
added at known concentration(s).
3
F1877-24
much as possible.When there is uncertainty regarding the
characteristics of particles produced from in vitro or bench
testing,particles from clinical studies(for example,retrievals)
can be used to enhance the clinical representativeness of testing
and its predictive power for characterizing potential biological
responses.
6.Interferences
6.1 Particles may form aggregates or agglomerates during
preparation and storage.These could result in an increase in
measured particle size.It is essential that care be taken to
resuspend particles prior to analysis and to note any effects of
the dispersant used.Details regarding resuspension of particles
prior to analysis are provided in AAMI TIR 42.
6.2 Particles from wear tests or harvested from tissues may
contain a mixture of materials.In order to determine all
potential sources of particle generation,appropriate methods
should be utilized to determine the chemical composition of the
entire particulate sample.Care should be taken when separat-
ing the particles and analyzing the physical characteristics and
chemical composition of the isolated particles.
6.3 Many automated particle counters operate on the as-
sumption that the particles are spherical.These methods may
not be appropriate for nonspherical debris.Additional methods
should be used to verify size,shape,and quantity using
methods that take aspect ratio into consideration,for example,
SEM or TEM image analysis.
7.Apparatus
7.1 Scanning Electron Microscope(SEM)(see Practice
E766,ISO 19749):
7.1.1 Standard SEM equipment can be utilized for many
studies.In special instances,such as with polymeric particles,
a low acceleration voltage(1 to 2 kV)machine with a high
brightness electron source,such as a field emission tip,may be
utilized.
7.1.2 Elemental analysis may be accomplished with an
energy dispersive X-ray spectrometer(EDS)with SEM.
7.1.3 Fourier transform infrared spectroscopy(FTIR)and
Raman spectroscopy are robust analytical methods to identify
the chemical composition(that is,chemical compounds pres-
ent)of bulk or individual particles.
7.2 Transmission Electron Microscopy (TEM):
7.2.1 TEM equipment can be typically used for the analysis
of nanoparticles,although SEM with a field emission tip has
also been successfully used to characterize particles as small as
50 to 100 nm.
7.2.2 Elemental analysis may be accomplished with an
energy dispersive X-ray spectrometer(EDS)coupled with
TEM.
7.3 Optical Microscope—An optical microscope operating
in the transmission mode may be utilized.Dark field illumina-
tion may enhance visualization of some particles.Polarized
light will facilitate identification of semicrystalline polymeric
materials.
7.4 Automatic Particle Counters(see Practice F660):
4.1 When evaluating particles generated from medical
devices,the final device within their designated commercial
packaging,including sterilization as applicable,should be used
in all evaluations.
4.2 Particles produced by implant wear in vivo in animal or
clinical studies are harvested from tissues after digestion
utilizing methods such as those in Practice F561.
4.3 Particles generated in vitro or obtained from commercial
sources are used as received,or after digestion if they were
generated in protein solutions,and further separation if there
are signs of aggregation.Harvesting and further processing
should maintain the original range and physical and chemical
characteristics of particles from in vivo or in vitro studies.For
routine analysis,the particles are characterized by the terms of
morphology and by size using Feret diameters and equivalent
circle diameter(ECD).For more detailed studies,several
methods that may be utilized for numerically characterizing
their dimensions and size distribution are described.
5.Significance and Use
5.1 The biological response to materials in the form of small
particles,as produced from implant wear,abrasion,or erosion,
often is significantly different from that to the same materials
in bulk form (that is,an implant component).Additionally,the
morphology(for example,size and shape,surface
characteristics),volume distribution,and species of these
particles are major determinants of device-related biological
responses;therefore,this practice provides standardized no-
menclature for describing particles.Such a unified nomencla-
ture will be of value in interpretation of biological tests of
responses to particles,in that it will facilitate separation of
biological responses per different particle characteristics such
as size,shape,and volume.
5.2 Particles released due to wear from implants in vivo may
result in an adverse biological response which will affect the
long-term survival of the device.Characterization of such
particles will provide valuable information regarding the safety
and effectiveness of device designs or methods of processing
components and the mechanisms of wear.
5.3 The morphology of particles produced in laboratory
tests of wear and abrasion often is affected by the test
conditions,such as the magnitude and rate of load application,
device configuration,and test environment.Comparison of the
morphology,size,and quantity of particles produced in vitro
with those produced in vivo will provide valuable information
regarding the degree to which the method simulates the in vivo
condition being modeled.
5.4 Particles harvested from particle-release studies(for
example,cell culture experiments,third body wear simulation)
that are to be used for testing should be representative of the
entire spectrum of possible particles produced from clinical use
of the device/material under review(for example,due to wear,
abrasion,or erosion).Therefore,efforts should be made to
ensure that the particles for testing were produced from in vivo
/in vitro studies that mimicked the clinical use conditions as
4.Summary of Practice

标签: #ASTM #颗粒 #表征

摘要:

1ThisinternationalstandardwasdevelopedinaccordancewithinternationallyrecognizedprinciplesonstandardizationestablishedintheDecisiononPrinciplesfortheDevelopmentofInternationalStandards,GuidesandRecommendationsissuedbytheWorldTradeOrganizationTechnicalBarrierstoTrade(TBT)Committee.Designation:F1877-24...

展开>> 收起<<
ASTM F1877-24 Characterization of Particles.pdf

共16页,预览16页

还剩页未读, 继续阅读

声明:如果您的权利被侵害,请联系我们的进行举报。
作者:冒牌货 分类:法规规范 价格:100质量币 属性:16 页 大小:1.12MB 格式:PDF 时间:2026-01-05
/ 16
客服
关注