Designation:F2504-05(Reapproved 2022)
Standard Practice for
Describing System Output of Implantable Middle Ear
Hearing Devices¹
This standard is issued under the fixed designation F2504;the number immediately following the designation indicates the year of
original adoption or,in the case of revision,the year of last revision.A number in parentheses indicates the year of last reapproval.A
superscript epsilon(ε)indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice defines means for describing system per-
formance(ex vivo)and,in particular,system output of an
implantable middle ear hearing device (IMEHD)by measuring
a physical quantity that is relevant to the insertion gain and
output level of the IMEHD when implanted in the patient.
1.2 This practice is similar to headphone calibration on an
artificial ear in which the sound pressure level (in decibel
sound pressure level (SPL))measured in the artificial ear can
be converted to patient hearing level (in decibel hearing level
(HL)using a known transfer function,as defined by ANSI 3.7.
These measurements can then be used to predict system
parameters relevant for patient benefit such as functional gain,
maximum output,and variability.Measurements defined in this
practice should be useful for patients,clinicians,
manufacturers,investigators,and regulatory agencies in mak-
ing comparative evaluations of IMEHDs.
1.3 The values given in SI units are to be considered the
standard.
1.4 This standard does not purport to address all of the
safety concerns,if any,associated with its use.It is the
responsibility of the user of this standard to establish appro-
priate safety,health,and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accor-
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
Development of International Standards,Guides and Recom-
mendations issued by the World Trade Organization Technical
Barriers to Trade(TBT)Committee.
¹This practice is under the jurisdiction of ASTM Committee F04 on Medical and
Surgical Materials and Devices and is the direct responsibility of Subcommittee
F04.37 on Implantable Hearing Devices(IHDs).
Current edition approved Oct.1,2022.Published October 2022.Originally
approved in 2005.Last previous edition approved in 2014 as F2504-05(2014).
DOI:10.1520/F2504-05R22.
2.Referenced Documents
2.1 ANSI Standards:²
ANSI 3.6 Specification for Audiometers
ANSI 3.7 Method for Coupler Calibration of Earphones
ANSI 3.22 Specification of Hearing Aid Characteristics
3.Terminology
3.1 Refer to the block diagram of Fig.1 for a clarification of
the mathematical notations used in this section.
3.2 In the following definitions,these symbols are used for
physical quantities:
3.2.1 E=electrical drive signal(voltage or current)
3.2.2 p=sound pressure
3.2.3 v=vibration velocity
3.3 All transfer functions are denoted by the symbol H,with
the following subscripts indicative of the type of transfer
function:
3.3.1 A=IMEHD-aided
3.3.2 E=electrical
3.3.3 H=hearing level
3.3.4 S=sound field sound pressure
3.3.5 T=tympanic membrane (ear drum)sound pressure
3.3.6 U=unimplanted
3.3.7 V=vibration of stapes
3.4 Definitions:
3.4.1 coupling,n— points and methods of attachment.
3.4.2 displacement,n—integral of velocity measured in
nanometres.
3.4.3 ear-canal sound pressure,pr n—sound pressure pro-
duced in the ear canal,at the tympanic membrane,by a sound
field stimulus,specified in units of pascals.
3.4.4 equivalent hearing level, LH,n—ratio of an equivalent
sound pressure,Po,relative to the sound field pressure,
2 Available from American National Standards Institute(ANSI),25 W.43rd St.,
4th Floor,New York,NY 10036,http://www.ansi.org.
Copyright ◎ASTM Intermational,100 Barr Harbor Drive,PO Box C700,West Conshohocken,PA 19428-2959.United States
F2504-05(2022)
UNIMPLANTED MIDDLE EAR
SOUND
FIELD p
EXTERNAL
EAR PT
TM &
OSSICLES
Vu
STAPES&
INNER
EAR
IMEHDAIDED MIDDLE EAR
OUTPUT
TRANSDUCER
SOUND
FIELD
STAPES&
INNER
EAR
VA
E
FIG.1 Signal Flow in the Unimplanted and IMEHD-Aided Middle Ear
PRETSPL,at 0°incidence that is just detectable monaurally by a
normally hearing individual,as defined in ANSI S3.6,Table 9,
expressed in decibels:LH=20-l1og10PoPRETSPL).
3.4.5 equivalent sound pressure,Pon— unimplanted input
sound field pressure needed to produce a stapes velocity equal
to that produced by a specified IMEHD input in the IMEHD-
aided condition:Po=E·HEs ·
3.4.5.1 Discussion—The equivalent sound pressure is the
product of the equivalent sound pressure transfer function,HES,
and the IMEHD output transducer electrical input E:Po=E ·
HEs. The equivalent sound pressure can be expressed as
equivalent sound pressure level in units of decibels, SPLeq
calculated as 20-1og10po/2·10-⁵Pa).
3.4.6 equivalent sound pressure level,Lo,n—logarithmic
representation of equivalent sound pressure,Lo=20-1og10PQ).
3.4.7 hearing level(HL),L,n—ratio of the input sound field
pressure,ps,relative to the sound field pressure PRETSPLat 0°
incidence that is just detectable monaurally by a normally
hearing individual,as defined in ANSI S3.6,Table 9,expressed
in decibels as:L=20·log10(Ps/PRETSPL).
3.4.8 IMEHD electrical input at threshold Erhreshold
n—electrical input to the IMEHD output transducer at thresh-
old of audibility.
3.4.9 IMEHD harmonic distortion, n—harmonic distortion
of the stapes velocity IMEHD-aided analogous to ANSI S3.22,
Section 6.11S,from sinusoidal inputs ofthe frequencies 500,
800,and 1600 Hz;input levels shall be Emax- 20 dB.
3.4.10 IMEHD output transducer; n—electromechanical
output transducer of the IMEHD.
3.4.11 IMEHD output transducerfrequency range,n—using
the equivalent sound pressure transfer function,HEs,draw a
horizontal line at the average for 1000,1600,and 2500 Hz,
then subtract 20 dB,or divide by 10;the lower and the upper
bounds ofthe frequency response range are where the average
line crosses the transfer function curve.
3.4.12 IMEHD output transducer input,E,n—electrical
input to the IMEHD output transducer,specified in volts or
amperes,as appropriate for the particular device.
3.4.13 IMEHD system frequency range,n—using the inser-
tion gain transfer function(velocity),Hv,draw a horizontal
line at the average for 1000,1600,and 2500 Hz,then subtract
20 dB,or divide by 10;the lower and the upper bounds ofthe
frequency response range are where the average line crosses
the transfer function curve.
3.4.14 input sound field pressure, Ps,n—sound stimulus
measured in the free field and presented to the listener in either
the IMEHD-aided or unimplanted condition,specified in units
of pascals.
3.4.15 maximum electrical transducer input,Emax
n—maximum electrical output of the sound signal processor,
specified as peak-to-peak or root mean square value,specified
in volts or amperes,as appropriate for the particular device.
3.4.16 maximum equivalent sound pressure,PE,max
n—equivalent sound pressure that corresponds to the maximum
electrical output Emax ofthe implant electronics,PEmax=Emax'
HES·
3.4.17 maximum equivalent sound pressure level,LE,max
n—logarithmic representation of the maximum equivalent
sound pressure LE,max=20-log10(PEmax/2·10-⁵Pa).
3.4.18 soundpressure at threshold,Pthreshold n—stimulus
sound field pressure at the threshold of audibility.
3.4.19 stapes velocity(IMEHD-aided),VA, n—translational
velocity of the stapes when driven by the IMEHD output
transducer,specified in units of mm/s.
3.4.20 stapes velocity(unimplanted) ,Vu,n—translational
velocity of the stapes when driven by sound input to the middle
ear specified in units of mm/s.
Transfer Function
3.4.21 acousto-electric transfer function,HsE, n—electrical
input to the IMEHD output transducer E produced by a sound
field,divided by the input sound field pressure ps:HsE=E/ps.
3.4.21.1 Discussion—HsE will depend on the particular gain
settings used,for example,full-on gain or minimal gain.The
gain should be reported whenever that transfer function is used.
3.4.22 acousto-vibrational transfer function(IMEHD
aided),HsvA—stapes velocity(IMEHD aided)divided by the
input sound field pressure:HsvA=VAps.
3.4.22.1 Discussion—This quantity can be measured di-
rectly or computed from the product ofthe electro-vibrational
F2504-05(2022)
3
transfer function,HEv,and the acousto-electric transfer
function,HsE,measured in the IMEHD-aided condition:HsvA
=VA/ps.
3.4.23 acousto-vibrational transfer function(unimplanted),
Hsvu, n—stapes velocity(unimplanted)when driven by the
input sound field,divided by the input sound field pressure:
Hsvu =Vu/ps.
3.4.23.1 Discussion—This quantity can be measured di-
rectly or computed from the product of the middle-ear transfer
function,HTv,and the ear-canal transfer function,Hsr,mea-
sured in the unimplanted condition:Hsvu=vU/Ps=HsrHTv.
3.4.24 ear-canal pressure transfer function,Hsz, n—ear
canal sound pressure,Pr,produced by the input sound field
pressure,ps,in the unimplanted case,divided by that input
sound field pressure:Hsr=Prps;this quantity is unitless(1,
2)³
.
3.4.25 electro-vibrational transferfunction,HEv ,n—stapes
velocity(IMEHD-aided)when driven by the IMEHD output
transducer,divided by the transducer input: HEv=vA/E.
3.4.26 equivalent sound pressure transfer function,HES,
n—unimplanted sound field pressure needed to produce a
stapes velocity equivalent to that produced by an electrical
IMEHD input in the IMEHD-aided condition,divided by the
IMEHD input.
3.4.26.1 Discussion—If the electrical IMEHD input pro-
duces a linear change in stapes velocity with a change in input
electrical stimulus,the equivalent sound pressure transfer
function,HEs,can be computed as the quotient between the
vibro-electric transfer function (IMEHD-aided),HEv,and the
vibro-acoustic transfer function (unimplanted), Hsvu:HEs=
(v/E)/(v/ps)= HEVHsvu.
3.4.27 insertion gain transfer function(sound field),Hss,
n—ratio of the equivalent sound pressure produced in the
IMEHD-aided case with a given electrical input to the IMEHD
output transducer and the input sound field pressure used as
input in the IMEHD-aided case required to produce the same
IMEHD output transducer electrical input:Hss=Pe/ps; this
ratio is unitless.
3.4.27.1 Discussion—With a linear sound signal processor,
the insertion gain(sound field)can be computed from the
product of the equivalent sound pressure transfer function,HES,
and the electro-acoustic transfer function, HsE:Hss=Pe/Ps=
Hse·Hes.Hss willdepend on the particular gain settings used,
for example,full-on gain or minimal gain.The gain should be
reported whenever that transfer function is used.
3.4.28 insertion gain transfer function(velocity),HVv
n—ratio of the stapes velocity (IMEHD-aided)and the stapes
velocity(unimplanted)produced by a given input sound field:
Hvv =VAVu;the ratio is unitless and can be expressed in
decibels as 20-log10(HV).
3.4.28.1 Discussion—With a linear sound signal processor
and IMEHD,that is,a processor whose electrical output E is
proportional to the input sound field pressure,ps,and an
IMEHD whose vibrational output is proportional to its electri-
3The boldface numbers in parentheses refer to the list of references at the end of
this standard.
cal output,the insertion gain(sound field),Hss,willequal the
insertion gain transfer function (velocity),Hv.
3.4.29 maximum insertion gain transfer function(sound
field),Hss,maxn n—maximum insertion gain transfer function
(sound field)that can be achieved with the implant electronics.
3.4.30 middle-ear transferfunction, HT,n—stapes velocity
(unimplanted)produced by an ear-canal sound pressure,di-
vided by the ear-canal sound pressure,in units of mm/s/Pa:
HTv=vUPr ·
3.5 Acronyms:
3.5.1 IHD—implantable hearing device
3.5.2 IMEHD—implantable middle-ear hearing device
3.5.3 LDV—laser Doppler vibrometry
3.5.4 SPL—sound pressure level
4.Summary of Practice
4.1 This practice involves the use of human temporal bones
and laser Doppler interferometry measurements of middle ear
structures velocities,to test for the ex-vivo performances of
IMEHD.Once a procedure for measuring system output has
been defined,several characteristics of the IMEHD can be
specified.Detailed instructions for measuring and reporting
these characteristics are given below.The important character-
istics are:
4.1.1 For Transducers:
4.1.1.1 Equivalent sound pressure transfer function,
4.1.1.2 IMEHD output transducer frequency range,and
4.1.1.3 IMEHD harmonic distortion.
4.1.2 For the System:
4.1.2.1 Maximum insertion gain transfer function(sound
field)(see full-on gain in ANSI S3.22,paragraph 3.7),
4.1.2.2 Maximum equivalent sound pressure level(see
OSPL90 in ANSI S3.22,paragraph 3.5),and
4.1.2.3 IMEHD system frequency range.
5.Significance and Use
5.1 IMEHDs are alternatives to air conduction hearing aids.
They are similar to air conduction hearing aids in that they
process incoming sound by applying frequency shaping and
compression to create an analog,vibratory audio frequency
output.IMEHDs differ from hearing aids in that they do not
create an airborne acoustical output signal with an electroa-
coustical output transducer in the external ear canal,but rather
a mechanical stimulation that results in the vibration of the
cochlear fluid.Therefore,the IMEHD output signal is not
readily accessible after implantation in the way hearing aid
output is accessible with real-ear probe microphone measure-
ments.Different devices will use different methods of coupling
to the ossicular chain or cochlea.This makes it difficult to
design a uniform model of the middle ear in the way the 2-cm³
coupler is used as a model of the external ear canal with
conventional hearing aids.
5.2 This practice provides uniformity of data collection
practices,thus allowing IMEHD in vitro performances to be
evaluated and readily compared.Once clinical data are
摘要:

1ThisinternationalstandardwasdevelopedinaccordancewithinternationallyrecognizedprinciplesonstandardizationestablishedintheDecisiononPrinciplesfortheDevelopmentofInternationalStandards,GuidesandRecommendationsissuedbytheWorldTradeOrganizationTechnicalBarrierstoTrade(TBT)Committee.Designation:F2504-05...